beautypg.com

20 current calibration for lpal, Lpal – AMETEK ReFlex User Manual

Page 332

background image

ReFlex Power™

Active Loads

M380056-01 Rev J

7-57

7.6.20

C

URRENT

C

ALIBRATION FOR

LPAL

This procedure calibrates the current for the high range. This is 5-point
calibration. Make sure that the shunt and the DC current meter are
connected between the active load and DC power supply.

This procedure requires maximum voltage 5.0 V and maximum current 15.0
Amps.

1. Set the DC input voltage to 5 Volts

2. Turn on the active load

INP:STATE 1

3. Set the DAC count to 5% of its full range

CAL:INP:CURR:COUNTS 5700

4. Wait for the DC current to stabilize. After the current reading is

stabilized, enter the current using following command

CAL:INP:CURR:FIVEPOINT1

5. Set the DAC count to 27.5% of its full range

CAL:INP:CURR:COUNTS 16000

6. Wait for the DC current to stabilize. After the current reading is

stabilized, enter the current using following command

CAL:INP:CURR:FIVEPOINT2

This manual is related to the following products: