20 current calibration for lpal, Lpal – AMETEK ReFlex User Manual
Page 332
ReFlex Power™
Active Loads
M380056-01 Rev J
7-57
7.6.20
C
URRENT
C
ALIBRATION FOR
LPAL
This procedure calibrates the current for the high range. This is 5-point
calibration. Make sure that the shunt and the DC current meter are
connected between the active load and DC power supply.
This procedure requires maximum voltage 5.0 V and maximum current 15.0
Amps.
1. Set the DC input voltage to 5 Volts
2. Turn on the active load
INP
3. Set the DAC count to 5% of its full range
CAL
4. Wait for the DC current to stabilize. After the current reading is
stabilized, enter the current using following command
CAL
5. Set the DAC count to 27.5% of its full range
CAL
6. Wait for the DC current to stabilize. After the current reading is
stabilized, enter the current using following command
CAL