When an DC power supply module is first powered on, the following
parameters are set to factory default values:

Output configuration: standalone

Output voltage: 0V

Output current: 0A

Overvoltage protection (OVP): 110% of full-scale voltage;
programmable mode

Undervoltage protection (UVP): 0V

Overcurrent protection (OCP): 120% of full-scale current;
programmable mode

Output state: off

Output isolation and remote sense relays: open

R

ESET

C

ONDITIONS

If a module is reset with the SCPI command, *RST[n], (where n is the module
address), the following default conditions are set:

Output configuration: unchanged

Output voltage: 0V

Output current: 0A

Overvoltage protection (OVP): 110% of full-scale voltage;
programmable mode

Undervoltage protection (UVP): 0V

Overcurrent protection (OCP): 120% of full-scale current;
programmable mode

Output state: off

Output isolation and remote sense relays: open

All status reporting data structures are cleared; enable-masks
are not cleared.

6.5.6

I

NITIAL

F

UNCTIONAL

T

ESTS

P

OWER

-O

N

C

HECK

1. Ensure that the AC/DC input service is disconnected.

2. Install the interface cable between the host controller and the

ReFlex Power™ Controller module.

3. Connect an appropriate output load to the DC power supply

module.