beautypg.com

AMETEK ReFlex User Manual

Page 336

background image

ReFlex Power™

Active Loads

M380056-01 Rev J

7-61

7. Send the calibration command to calculate the gain and offset of the

low range current calibration

CAL:INP:CURR:CALC

8. Send the system error query command to make sure calibration was

successful.

SYST:ERR?

9. If there is no error returned at the system error query command then

save the calibration result with the following command

CAL:UNLOCK “6867”

CAL:STORE

CAL:LOCK

10. Reset the unit

*RST