AC Power Supplies

ReFlex Power™

4-80

M380056-01 Rev K

14. Set the output frequency to 250Hz

SOUR:FREQ 250

15. Wait for AC current to stabilize. Measure the output current using the AC current

meter. Enter the measured current using following command

CAL:OUTP:FREQ:CURR2

16. Set the output frequency to 440Hz

SOUR:FREQ 440

17. Wait for AC current to stabilize. Measure the output current using the AC current

meter. Enter the measured current using following command

CAL:OUTP:FREQ:CURR3

18. Set the output frequency to 630Hz

SOUR:FREQ 630

19. Wait for AC current to stabilize. Measure the output current using the AC current

meter. Enter the measured current using following command

CAL:OUTP:FREQ:CURR4

20. Set the output frequency to 820Hz

SOUR:FREQ 820

21. Wait for AC current to stabilize. Measure the output current using the AC current

meter. Enter the measured current using following command

CAL:OUTP:FREQ:CURR5

22. Set the output frequency to 1010Hz

SOUR:FREQ 1010

23. Wait for AC current to stabilize. Measure the output current using the AC current

meter. Enter the measured current using following command

CAL:OUTP:FREQ:CURR6

24. Set the output frequency to 1200Hz

SOUR:FREQ 1200

25. Wait for AC current to stabilize. Measure the output current using the AC current

meter. Enter the measured current using following command

CAL:OUTP:FREQ:CURR7

26. Perform the voltage calibration using following command