AMETEK ReFlex User Manual
Page 165

AC Power Supplies
ReFlex Power™
4-80
M380056-01 Rev K
14. Set the output frequency to 250Hz
SOUR
15. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current using following command
CAL
16. Set the output frequency to 440Hz
SOUR
17. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current using following command
CAL
18. Set the output frequency to 630Hz
SOUR
19. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current using following command
CAL
20. Set the output frequency to 820Hz
SOUR
21. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current using following command
CAL
22. Set the output frequency to 1010Hz
SOUR
23. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current using following command
CAL
24. Set the output frequency to 1200Hz
SOUR
25. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current using following command
CAL
26. Perform the voltage calibration using following command