CAL:OUTP:FREQ:CALC

27. Send the system error query command to make sure calibration was successful.

SYST:ERR?

28. If there is no error returned at the system error query command then save the

calibration result with the following commands

CAL:UNLOCK “6867”
CAL:STORE
CAL:LOCK

29. Reset the unit

*RST

This manual is related to the following products: