AMETEK ReFlex User Manual
Page 140
ReFlex Power™
AC Power Supplies
M380056-01 Rev K
4-55
4.6.9
V
OLTAGE
ADC
O
FFSET
C
ALIBRATION FOR
V
OLTAGE
R
ANGE
H
IGH
The following procedure calibrates offset of the voltage ADC for the voltage
range high.
1. Ensure that no load is connected to the output.
2. Issue clear and reset commands:
*CLS
*RST
3. Set the voltage range of the power supply to high
SOUR
4. Perform the voltage ADC offset calibration
CAL
5. Send the system error query command to make sure calibration was successful.
SYST:ERR?
6. If there is no error returned at the system error query command then save the
calibration result with the following command
CAL
CAL
CAL
7. Optional Commands:
To query the voltage ADC offset:
CAL
8. Reset the unit
*RST