beautypg.com

6 regulatory compliance, 4 minimum requirements, Minimum requirements – AMETEK ReFlex User Manual

Page 34

background image

System Overview

ReFlex Power™

1-12

M380056-01 Rev J

Calibration Temperature Range:

Performance specifications are valid following calibration at
25

°C, +/- 5°C

N

ON

-

OPERATING

E

NVIRONMENTAL

C

ONDITIONS

(T

RANSPORTABILITY

):

Temperature:

-40 to 71

°C

Altitude:

0 to 50,000 feet

Humidity:

95% relative humidity non-condensing.

Shock:

Class 3 MIL-PRF-28800F

Vibration:

Class 3 MIL-PRF-28800F

1.3.6

R

EGULATORY

C

OMPLIANCE

EMC compliance is per EN 61326:1997 and safety compliance is per IEC

61010-1; satisfies MIL-PRF-28800F

Test Environment

CISPR 16-1

Conducted Susceptibility

IEC 1000-4-6, Conducted RF
IEC 1000-4-2, ESD
IEC 10004-4, Burst
IEC 61000-4-5, Surge
IEC 61000-4-11, Input Power Interruptions

Conducted Emission

CISPR 11, CLASS A

Radiated Susceptibility

IEC 1000-4-3, EM FIELD

Radiated Emission

CISPR 11, CLASS A

1.4

MINIMUM REQUIREMENTS

The minimum requirements to operate a ReFlex Power™ system are:

Mainframe

Controller module

One or more power modules

This manual is related to the following products: