AMETEK ReFlex User Manual
Page 333
Active Loads
ReFlex Power™
7-58
M380056-01 Rev J
7. Set the DAC count to 50.0% of its full range
CAL
8. Wait for the DC current to stabilize. After the current reading is
stabilized, enter the current using following command
CAL
9. Set the DAC count to 72.5% of its full range
CAL
10. Wait for the DC current to stabilize. After the current reading is
stabilized, enter the current using following command
CAL
11. Set the DAC count to 95% of its full range
CAL
12. Wait for the DC current to stabilize. After the current reading is
stabilized, enter the current using following command
CAL
13. Set the DAC to zero counts
CAL