4 calpot calibration for voltage range low – AMETEK ReFlex User Manual

Page 131

background image

AC Power Supplies

ReFlex Power™

4-46

M380056-01 Rev K

Figure 4-14. Calibration Instrumentation Setup

4.6.4

C

AL

P

OT

C

ALIBRATION FOR

V

OLTAGE

R

ANGE

L

OW

The following procedure calibrates cal-pot for the voltage range low. This
calibration is done with the AC current meter connected to the output lead of the
AC power supply.

1. Connect the AC current meter to the output lead of the power supply. Ensure that

no load is connected to the output.

2. Issue clear and reset commands:

*CLS
*RST

3. Start the internal calibration of the offset pot

CAL:CALPOTS:OFFSET:CALC