6 calibration, 1 scope – AMETEK ReFlex User Manual
Page 129
AC Power Supplies
ReFlex Power™
4-44
M380056-01 Rev K
6. Turn on the isolation and remote sense relays, turn on the output
converters, and program full scale voltage and current with the following
commands. The parameter,
current of the particular model under test. The parameter, [1000+n1], is
the sum of 1000 and the slot number of the master module. For example,
if the master module is located in slot 4, the parameter value would be
1001.
OUTP[1000+n1]:STATE 1
SOUR[1000+n1]:CURR
SOUR[1000+n1]:VOLT
7. Measure the output voltage and current with the external instruments.
8. Ensure that the actual values of the output parameters are within
specifications of the programmed values. The default frequency is 60 Hz.
9. Query the values of the output parameters with the following commands:
MEAS[1000+n1]:VOLT?
MEAS[1000+n1]:CURR?
MEAS[1000+N1]:FREQ?
10. Ensure that the readback values of the output parameters are within
specifications of the actual values.
11. Turn off the output with the following command:
*RST[n]
4.6
CALIBRATION
4.6.1
S
COPE
Procedures are provided in the following sections for calibration of the
ReFlex
Power
TM
AC Power supply module: ReFlex Power
TM
ACPS 1.2KHz module, P/N:
RFP-A301K-875-XXXX, and ReFlex Power
TM
ACPS 5.0KHz module, P/N: RFP-
A305K-875-XXXX. Calibration is easily performed through the Controller module
interface with SCPI commands, requiring only 2 meters and resistive load.