Active Loads

ReFlex Power™

7-54

M380056-01 Rev J

8. Set the second power supply (Power supply connected to IAI Board)

voltage to 10.00 volts.

9. After the DMM voltage stabilizes enter the voltage of the DMM

(Connected to second power supply or Power supply connected to
IAI Board.)

Note: This voltage should be equal to 10.0 Volts.

CAL:INP:VOLT2

10. Send the command to calculate the gain and offset of the IAI board.

CAL:MEAS:OFFSET:CALC

11. Set the active load back to the SCPI mode

INP:CURRent:EXTernal 0

12. Send the system error query command to make sure calibration was

successful.

SYST:ERR?

13. If there is no error returned at the system error query command then

save the calibration result with the following command

CAL:UNLOCK “6867”

CAL:STORE

CAL:LOCK

14. Reset the unit

*RST