Active Loads
ReFlex Power™
7-54
M380056-01 Rev J
8. Set the second power supply (Power supply connected to IAI Board)
voltage to 10.00 volts.
9. After the DMM voltage stabilizes enter the voltage of the DMM
(Connected to second power supply or Power supply connected to
IAI Board.)
Note: This voltage should be equal to 10.0 Volts.
CAL
10. Send the command to calculate the gain and offset of the IAI board.
CAL
11. Set the active load back to the SCPI mode
INP
12. Send the system error query command to make sure calibration was
successful.
SYST:ERR?
13. If there is no error returned at the system error query command then
save the calibration result with the following command
CAL
CAL
CAL
14. Reset the unit