ReFlex Power™

AC Power Supplies

M380056-01 Rev K

4-47

4. Start the calibration of the gain pot

CAL:CALPOTS:GAIN:START

5. Set the load box to approximately 6.5Amps. Wait for the current to stabilize then

make the AC current measurement using the meter connected to the power
supplies output lead.

CAL:CALPOTS:GAIN:CURR

6. Finish the gain pot calibration

CAL:CALPOTS:GAIN:CALC

7. Send the system error query command to make sure calibration was successful.

SYST:ERR?

8. If there is no error returned at the system error query command then save the

calibration result with the following commands

CAL:UNLOCK “6867”
CAL:STORE
CAL:LOCK

9. Optional Commands:

T

o query the offset pot set value

:

CAL:CALPOTS:OFFSET?

To query the gain pot set value:

CAL:CALPOTS:GAIN?

10. Reset the unit

*RST