ReFlex Power™
AC Power Supplies
M380056-01 Rev K
4-47
4. Start the calibration of the gain pot
CAL
5. Set the load box to approximately 6.5Amps. Wait for the current to stabilize then
make the AC current measurement using the meter connected to the power
supplies output lead.
CAL
6. Finish the gain pot calibration
CAL
7. Send the system error query command to make sure calibration was successful.
SYST:ERR?
8. If there is no error returned at the system error query command then save the
calibration result with the following commands
CAL
CAL
CAL
9. Optional Commands:
T
o query the offset pot set value
:
CAL
To query the gain pot set value:
CAL
10. Reset the unit
*RST