LPAL

V

OLTAGE

C

ALIBRATION

P

ART

II

OF

V

This procedure calibrates the voltage for voltage gain 4 and current range
high. Make sure that the shunt and the DC voltmeter are connected
between the positive and negative lead of the active load. Also connect the
sense positive-to-positive lead and sense negative-to-negative lead.

This procedure requires maximum voltage 109 V and maximum current 3.0
Amps.

1. Turn the active load on

INP:STATE 1

2. Set the conductance count to 0x6666

CAL:INP:COND:COUNTS 0x6666

3. Set the input DC voltage to 27 Volts. After the input DC voltage

stabilizes enter the first exact voltage using following command

CAL:INP:VOLT 1

4. Set the input DC voltage to 109 Volts. After the input DC voltage

stabilizes enter the second exact voltage using following command

CAL:INP:VOLT 2

5. Send the calibration command to calculate the gain and offset:

CAL:MEAS:OFFSET:CALC

6. Send the system error query command to make sure calibration was

successful.

SYST:ERR?