AMETEK ReFlex User Manual
Page 149

AC Power Supplies
ReFlex Power™
4-64
M380056-01 Rev K
CAL
12. Program the output voltage to 50% of the full scale voltage
CAL
13. Wait for AC voltage to stabilize. Measure the output voltage using the AC
voltmeter. Enter the measured voltage using following command
CAL
14. Program the output voltage to 72.5% of the full scale voltage
CAL
15. Wait for AC voltage to stabilize. Measure the output voltage using the AC
voltmeter. Enter the measured voltage using following command
CAL
16. Program the output voltage to 95% of the full scale voltage
CAL
17. Wait for AC voltage to stabilize. Measure the output voltage using the AC
voltmeter. Enter the measured voltage using following command
CAL
18. Perform the voltage calibration using following command
CAL
19. Send the system error query command to make sure calibration was successful.
SYST:ERR?
20. If there is no error returned at the system error query command then save the
calibration result with the following command
CAL
CAL
CAL
21. Optional Commands:
Query the voltage calibration points: CAL
22. Reset the unit
*RST