Active Loads

ReFlex Power™

7-36

M380056-01 Rev J

7.6.6

HPAL

L

OW

R

ANGE

C

URRENT

C

ALIBRATION

The following procedure calibrates the input current for the current range
low. Make sure that the shunt and the DC current meter are connected
between the active load and DC power supply.

This procedure requires maximum voltage 5.0V and maximum current 1.0
Amps.

1. Set the DC input voltage to 5 Volts

2. Turn on the active load

INP:STATE 1

3. Set the DAC count to 20% of its full range

CAL:INP:CURR:COUNTS 16000

4. Wait for the DC current to stabilize. After the current reading is

stabilized, enter the current using following command

CAL:INP:CURR1

5. Set the DAC count to 80% of its full range

CAL:INP:CURR:COUNTS 53000

6. Wait for the DC current to stabilize. After the current reading is

stabilized, enter the current using following command

CAL:INP:CURR2

7. Send the calibration command to calculate the gain and offset of the

low range current calibration

CAL:INP:CURR:CALC

8. Send the system error query command to make sure calibration was

successful.

SYST:ERR?

9. If there is no error returned at the system error query command then

save the calibration result with the following command

CAL:UNLOCK “6867”

CAL:STORE

CAL:LOCK

10. Reset the unit

*RST