AMETEK ReFlex User Manual
Page 159
AC Power Supplies
ReFlex Power™
4-74
M380056-01 Rev K
15. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL
16. Set the output frequency to 1000Hz
SOUR
17. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL
18. Set the output frequency to 2000Hz:
SOUR
19. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL
20. Set the output frequency to 3000Hz:
SOUR
21. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL
22. Set the output frequency to 4000Hz:
SOUR
23. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL
24. Set the output frequency to 5000Hz:
SOUR
25. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
CAL
26. Perform the voltage calibration:
CAL
27. Send the system error query command to make sure calibration was successful:
SYST:ERR?