7.6.22

LPAL

V

OLTAGE

C

ALIBRATION

PART

I

OF

V

The following procedure calibrates the voltage for voltage gain 8 and
current range high. Make sure that the shunt and the DC voltmeter are
connected between the positive and negative lead of the active load. Also
connect the sense positive-to-positive lead and sense negative-to-negative
lead.

This procedure requires maximum voltage 55 V and maximum current
6.0Amps.

1. Turn the active load on

INP:STATE 1

2. Set the conductance count to 0x19999

CAL:INP:COND:COUNTS 0x19999

3. Set the input DC voltage to 13 Volts. After the input DC voltage

stabilizes enter the first exact voltage using following command

CAL:INP:VOLT 1

4. Set the input DC voltage to 54 Volts. After the input DC voltage

stabilizes enter the second exact voltage using following command

CAL:INP:VOLT 2

5. Send the calibration command to calculate the gain and offset:

CAL:MEAS:OFFSET:CALC

6. Send the system error query command to make sure calibration was

successful.

SYST:ERR?