ReFlex Power™

Active Loads

M380056-01 Rev J

7-75

7. After DC voltage stabilizes. Enter the voltage using SCPI

commands

CAL:INP:VOLT 2

8. After DC current stabilizes. Enter the current using SCPI commands

CAL:INP:CURR 2

9. Send the calibration command to calculate the gain and offset

CAL:INP:COND:CALC

10. Send the system error query command to make sure calibration was

successful.

SYST:ERR?

11. If there is no error returned at the system error query command then

save the calibration result with the following command

CAL:UNLOCK “6867”

CAL:STORE

CAL:LOCK

12. Reset the unit

*RST