ReFlex Power™
Active Loads
M380056-01 Rev J
7-75
7. After DC voltage stabilizes. Enter the voltage using SCPI
commands
CAL
8. After DC current stabilizes. Enter the current using SCPI commands
CAL
9. Send the calibration command to calculate the gain and offset
CAL
10. Send the system error query command to make sure calibration was
successful.
SYST:ERR?
11. If there is no error returned at the system error query command then
save the calibration result with the following command
CAL
CAL
CAL
12. Reset the unit
*RST