AMETEK ReFlex User Manual
Page 138
ReFlex Power™
AC Power Supplies
M380056-01 Rev K
4-53
14. Wait for the DC voltmeter to stabilize. Make an accurate measurement and then
enter the DC voltage using following command. User should enter the voltage
including sign (Ex: For Negative 4.15631 enter –4.15631)
CAL
15. Program to DAC to 0.5% above the middle range
CAL
16. Wait for the DC voltmeter to stabilize. Make an accurate measurement and then
enter the DC voltage using following command
CAL
17. Perform the DAC Calibration
CAL
18. Measure the DC output voltage, if the voltage is not 0.0 ± 10mVolts, read the
DAC offset value with CAL
returned (a Hex number) as a starting point, “walk” the DC voltage toward 0.0
volts by increasing the value if the DC value is negative, and decreasing the
value if it is positive. One count is
≈ 10mVolts. The command to change the
DAC offset value is CAL
19. Turn the DC correction loop back on
CAL
20. Turn off the unit
OUTP
21. Send the system error query command to make sure calibration was successful.
SYST:ERR?
22. If there is no error returned at the system error query command then save the
calibration result with the following command
CAL
CAL
CAL
23. Reset the unit
*RST