2 meas ph - general/hardware, General/hardware – Metrohm tiamo 2.3 Patch 1 User Manual
Page 953
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1
■■■■■■■■
937
Designation
Description
.LP.CAx
Calculated value x (1…3) for the last measuring point
on the measuring point list
.LP.EXx
External value x (1…3) for the last measuring point on
the measuring point list
.LP.MEA
Measured value for the last measuring point of the
measuring point list in the unit of the measured value
.LP.TEM
Temperature for the last measuring point of the meas-
uring point list in °C
.LP.TIM
Time in s until the last measuring point in the measur-
ing point list is reached
.MA.MEA
Maximum measured value in the unit of the measured
value
.MA.TEM
Temperature for the maximum measured value in °C
.MA.TIM
Time in s until the maximum measured value is reached
.MI.MEA
Minimum measured value in the unit of the measured
value
.MI.TEM
Temperature with minimum measured value in °C
.MI.TIM
Time for the minimum measured value in s
.MTE
Temperature measurement with sensor; 1 = on, 0 =
off)
.NMP
Number of measuring points in measuring point list
.SLO
Electrode slope of the sensor used for the command (in
%)
.STY
Type of stop with which the command was stopped: 1
= normal; 0 = manual or after error
5.6.4.3.2
MEAS pH - General/Hardware
Tab: Method
▶ MEAS pH ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Input
25 characters
The general parameters for the device, the sensor and the stirrer are
defined on this tab.