3 cal meas opt - measuring parameters – Metrohm tiamo 2.3 Patch 1 User Manual
Page 1145
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1
■■■■■■■■
1129
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Measuring parameters tab will be
used.
Stirrer
The section Stirrer is only displayed for Device type = 089 Photome-
ter.
Stirrer
Selection of the stirrer.
Selection
1 | off
Default value
off
Stirring rate
Setting the stirring rate.
Range
0 ... 15
Default value
3
Switch off automatically
on | off (Default value: on)
If this check box is activated, the stirrer will be switched off automatically
when the command has finished.
5.6.5.11.3
CAL MEAS Opt - Measuring parameters
Tab: Method
▶ CAL MEAS Opt ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Input
25 characters
The measuring parameters for the calibration are defined on this tab.
Parameters for the Avantes spectrometer
Wavelength
Wavelength at which the measured quantity (absorbance) is recorded.
Range
100.0 ... 2000.0 nm
Default value
400.0 nm
Blank value
Value by which the measured quantity (absorbance) is corrected. Here an
expression can also be entered through the formula editor (e.g. xx.EME