8 kfc - additional evaluations, Additional evaluations – Metrohm tiamo 2.3 Patch 1 User Manual
Page 856
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
840
■■■■■■■■
tiamo 2.3 Patch 1
Automatic start after sample addition
on | off (Default value: off)
If this option is enabled, the titration starts automatically when the com-
mand is in status COND OK and the Threshold value set is exceeded. In
this case the system assumes that the sample has been injected. This para-
meter is displayed only for 851 and 852.
Threshold value
Value the automatic start after the sample addition is carried out automat-
ically when it is exceeded.
851, 852
Range
0 ... 999 mV
Default value
50 mV
Only start titration by a start command from a SEND command
on | off (Default value: off)
If this option is activated, then the titration will not be started until the
Event message Start titration is sent to the command with a SEND
command.
Note
Conditioning must already be completed (= Condok) at the time the
SEND command is received, otherwise the command to start the titra-
tion will be ignored.
If this command is already started (e.g. via Automatic start after
sample addition), the SEND command will not have any effect.
Show measured value during conditioning
on | off (Default value: off)
If this option is enabled, then the measured value will be shown in the
subwindow Live display of the program part Workplace, in addition to
the drift and the time.
5.6.3.6.8
KFC - Additional evaluations
Tab: Method
▶ KFC ▶ Properties... ▶ Additional evaluations
Command name
Name of the command.
Input
25 characters