4 meas t/flow - evaluations, Evaluation – Metrohm tiamo 2.3 Patch 1 User Manual
Page 1019
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1
■■■■■■■■
1003
Range
0 ... 999999 s
Default value
600 s
Measuring interval
Time interval for entering a measuring point in the measuring point list.
Range
1 ... 999999.0 s
Default value
10 s
Stop measured value Temp.
Stops when the preset measured value has been reached since the start of
the measurement. off means no stop.
Range
50.0 ... 250.0 °C
Selection
off
Default value
off
5.6.4.8.4
MEAS T/Flow - Evaluations
Tab: Method
▶ MEAS T/Flow ▶ Properties... ▶ Evaluations
Command name
Name of the command.
Input
25 characters
The following additional methods for evaluation of measurement curves
can be activated and defined on this tab.
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is enabled, then the associated time for the minimum meas-
ured value will be interpolated from the measuring point list.
Threshold value
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.
Range
0.1 ... 20.0 °C/s
Default value
1.0 °C/s
Maximum evaluation
Maximum evaluation
on | off (Default value: off)
If this option is enabled, then the associated time for the maximum meas-
ured value will be interpolated from the measuring point list.