10 det u - volume window – Metrohm tiamo 2.3 Patch 1 User Manual
Page 527
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1
■■■■■■■■
511
Titrando, 855
Selection
first | greatest | last | ascending | descending
Default value
first
first
Only the first endpoint to be found will be recognized.
greatest
Only the endpoint with the greatest ERC value, i.e. the steepest jump,
will be recognized.
last
Only the last endpoint to be found will be recognized.
ascending
Only endpoints with a positive slope of the titration curve will be rec-
ognized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.
5.6.3.2.3.10
DET U - Volume window
Dialog window: Method
▶ DET U ▶ Properties... ▶ Potentiometric evalua-
tion
▶ Evaluation with volume window (ml) ▶ [New]/[Properties] ▶ Volume win-
dow #
Volume windows are defined sub-regions (windows) on the volume axis
for which different parameters for the potentiometric evaluation can be
defined. Only those endpoints which lie within these windows and addi-
tionally meet the parameters defined for each window will be recognized.
Only one endpoint will be recognized per window.
Lower limit
Lower limit for the volume window.
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
0.00000 mL
Upper limit
Upper limit for the volume window.
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
9999.99 mL