The conductivity
measured at a particular temperature is thus auto-
matically converted to the conductivity
of a freely selectable reference
temperature (usually 20 C or 25 C). The conversion takes place with
the aid of the temperature coefficients
with respect to the reference
temperature .
Conductivity at reference temperature
Temperature coefficient
Conductivity at reference temperature
Conductivity at measuring temperature
Measuring temperature
Reference temperature
Temperature coefficients with respect to the reference temperature
(unit: %/°C).
5.6.3.3.6.5
MET Cond - Stop conditions
Tab: Method
▶ MET Cond ▶ Properties… ▶ Stop condition
Command name
Name of the command.
Input
25 characters
Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop
the titration.
Stop volume
Stop when the entered volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.
Range
0.00000 ... 9999.99 mL
Default value
100.000 mL
Selection
off