5 special beaker – Metrohm tiamo 2.3 Patch 1 User Manual
Page 1476
6.12 Subwindow Rack data
■■■■■■■■■■■■■■■■■■■■■■
1460
■■■■■■■■
tiamo 2.3 Patch 1
Note
Special beakers should preferably be set at high rack positions so that
sample series can start from rack position 1. Rack positions which are
defined as special beakers are no longer available as sample positions. A
specific work position can be defined for each special beaker at Tower
1 and Tower 2 (if present). The shift, rinse and special positions of the
affected tower are taken from the general rack positions.
The following columns are shown in the table:
Special beaker
Number of the special beaker for the selected rack.
Rack position
Number of the rack position for the special beaker.
Work position Tower 1
Work position for the special beaker at Tower 1.
Work position Tower 2
Work position for the special beaker at Tower 2.
Beaker radius
Radius of the special beaker.
Beaker sensor
Shows whether and which beaker sensor is to be used for the special
beaker.
[Edit]
Opens the dialog window Special beaker # (see Chapter 6.12.3.5, page
1460) for editing the data of the selected special beaker.
6.12.3.5
Special beaker
Dialog window: Configuration
▶ Rack data ▶ [Edit] ▶ Properties... ▶ Rack
data
▶ Special beaker ▶ [Edit] ▶ Special beaker #
The parameters for the special beaker selected in the table of special beak-
ers can be defined in the dialog window Special beaker #.
Rack position
Number of rack position for selected special beaker.
Range
0 ... n (n is rack-dependent)