5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1118
■■■■■■■■
tiamo 2.3 Patch 1
Buffers
Number of buffers
Displaying the number of buffers (3), with which the electrode test is to be
carried out.
Buffer type
Selecting a predefined buffer series for the pH calibration. It is imperative
that the selected buffer series include the buffers 4, 7 and 9.
If these buffers are used, tiamo will recognize the particular buffer auto-
matically.
Selection
Metrohm | NIST | DIN | Fluka Basel | Mettler |
Merck CertiPUR 20°C | Merck CertiPUR 25°C |
Radiometer | Baker | Hamilton DURACAL | Pre-
cisa
Default value
Metrohm
Request for buffer exchange
on | off (Default value: on)
If this check box is activated, a request for manual buffer exchange is dis-
played after each time the calibration loop is run through. This parameter
must be disabled for automatic buffer exchange when calibrating with a
Sample Processor.
Electrode test
Electrode type
Selection of a predefined electrode type which is to be used to assess the
pH electrode.
Selection
Standard | Gel | Non-aqueous | 'Electrode type
name'
Default value
Standard
'Electrode type name'
Uses the own electrode type defined under this name in the configura-
tion under Extras
▶ Templates ▶ Electrode type.
Save calibration data
on | off (Default value: off)
If this check box is activated, then the Slope and pH(0) calibration data
determined at the time of the electrode test is stored in the configuration
with the sensor used.