4 det ph - titration parameters, Titration parameters – Metrohm tiamo 2.3 Patch 1 User Manual
Page 486
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
470
■■■■■■■■
tiamo 2.3 Patch 1
Dosing rate
Speed at which the start volume is added until the start slope is reached.
The maximum dosing rate depends on the cylinder volume of the
exchange unit or dosing unit used.
Range
0.01 ... 166.00 mL/min
Default value
5.00 mL/min
Selection
maximum
Pause
Pause
Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the addition of the start volume. The pause follows at
the end of all the start conditions.
Range
0 ... 999999 s
Default value
0 s
5.6.3.2.2.4
DET pH - Titration parameters
Tab: Method
▶ DET pH ▶ Properties... ▶ Titration parameters
Command name
Name of the command.
Input
25 characters
The parameters for the running of the titration are defined on this tab.
Titration rate
Titration rate
The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for the Measured value
acceptance and Dosing of increments of these sets are not displayed.
The User setting must be selected in order to be able to edit these param-
eters.
Selection
slow | optimal | fast | User
Default value
optimal
slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwan-
ted EPs.
optimal
Parameter set for all standard titrations; optimized for the most fre-
quent applications.