4 properties - limit values – Metrohm tiamo 2.3 Patch 1 User Manual
Page 1419
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
tiamo 2.3 Patch 1
■■■■■■■■
1403
Cancel determination
The running determination will be automatically canceled. The follow-
ing message must be confirmed with [OK].
6.7.3.4
Properties - Limit values
Tab: Configuration
▶ Sensors ▶ Edit ▶ Properties ▶ Limit values
Note
This tab is displayed only for sensors of the type pH electrode, ISE
electrode or conductivity sensor.
Slope monitoring (for pH and ISE electrodes only)
on | off (Default value: off)
If this option is activated, then a check will be made to see whether the
Slope determined during calibration is within the defined limits. If it is
outside them a corresponding message appears and you can select
whether you will nevertheless accept the value or whether the previous
value is to be retained.
Lower limit
Lower limit value for the slope.
for pH electrode
Range
-2.147E+9 ... 2.147E+9 %
Default value
96.0 %
for ISE (+1)
Range
-2.147E+9 ... 2.147E+9 mV
Default value
55.0 mV
for ISE (-1)
Range
-2.147E+9 ... 2.147E+9 mV
Default value
-61.0 mV
for ISE (+2)
Range
-2.147E+9 ... 2.147E+9 mV
Default value
25.0 mV
for ISE (-2)
Range
-2.147E+9 ... 2.147E+9 mV
Default value
-31.0 mV