Metrohm tiamo 2.3 Patch 1 User Manual
Page 883
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1
■■■■■■■■
867
855
Selection
1 | 2 | 3
Default value
1
Titrino
Selection
internal D0
736, 751, 799
Selection
internal D0 | external D1 | external D2
Default value
internal D0
Solution
Input of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent exchange or dosing units are
used, then a check will be made in the method sequence to see whether
the correct solution has been set on the selected dosing device and
whether the dosing device type is correct. With non-intelligent exchange
or dosing units, only the cylinder volume is checked. At the start of the
command, a check is made of the working life, the validity of the titer and
the GLP test interval for the selected solution.
Input
24 characters
Selection
'Solution name' | not defined
Default value
not defined
not defined
No tests will be carried out.
Tandem dosing
on | off (Default value: off)
If this option is enabled, the dosing is carried out without interruption
with a combination of two dosing devices so that the second dosing
device is dosing while the first one is being filled and vice-versa (see Chap-
ter 5.6.3.8.5, page 861). This parameter is displayed only for Titrandos
and 855.
Dosing device
Selection of the number of the dosing device (dosing or exchange unit) to
be used for the dosing. All the dosing device connections which are possi-
ble with the selected device type are displayed.
Titrando
Selection
1 | 2 | 3 | 4
Default value
2