10 set ipol - fixed endpoint evaluation, 6 set upol, 1 set upol - overview – Metrohm tiamo 2.3 Patch 1 User Manual
Page 776
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
760
■■■■■■■■
tiamo 2.3 Patch 1
5.6.3.4.5.10
SET Ipol - Fixed endpoint evaluation
Dialog window: Method
▶ SET Ipol ▶ Properties... ▶ Additional evalua-
tions
▶ Fixed endpoint evaluation ▶ [New]/[Properties] ▶ Fixed endpoint evalua-
tion #
Quantity
Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.
Selection
Measured value | Time | Volume
Default value
Measured value
Fix value
Value of the fix endpoint.
Measured value
Range
-2000.0 ... 2000.0 mV
Time
Range
0.0 ... 999999.9 s
Volume
Range
0.00000 ... 9999.99 mL
5.6.3.4.6
SET Upol
5.6.3.4.6.1
SET Upol - Overview
Dialog window: Method
▶ SET Upol ▶ Properties... ▶ SET Upol - 'Command name'
Command for Set endpoint titrations with amperometric measurement
(selectable polarization voltage).
Instruments
This command can be executed with the following devices:
Titrando: 808, 809, 835, 836, 841, 842, 852, 857, 888, 901, 902, 904,
905, 906, 907
Titrino: 702, 716, 718, 719, 720, 721, 736, 751, 758, 785, 794, 798,
799
Robotic Titrosampler: 855
Appearance
The command has the following appearance:
without conditioning
with conditioning