5 met upol, 1 met upol - overview, Met upol – Metrohm tiamo 2.3 Patch 1 User Manual
Page 663
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1
■■■■■■■■
647
5.6.3.3.5
MET Upol
5.6.3.3.5.1
MET Upol - Overview
Dialog window: Method
▶ MET Upol ▶ Properties... ▶ MET Upol - 'Command
name'
Command for Monotonic equivalence point titrations with ampero-
metric measurement (selectable polarization voltage).
Instruments
This command can be executed with the following devices.
Titrando: 808, 809, 835, 836, 852, 857, 888, 904, 905, 906, 907
Titrino: 702, 716, 721, 736, 751, 785, 794, 798, 799
Robotic Titrosampler: 855
Appearance
The command has the following appearance:
Parameters
The parameters for the command MET Upol are configured in the fol-
lowing 7 tabs:
■
General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.
■
Start conditions
Parameters for the measurement of the initial measured value and for
defining the conditions which have to be met at the start of the titra-
tion.
■
Titration parameters
Parameters for the run of the titration.
■
Stop conditions
Definitions of conditions which cause the stop of the titration.
■
Potentiometric evaluation
Parameters for the potentiometric evaluation of titration curves.
■
Additional evaluations
Definition of additional methods for the evaluation of the titration
curves.
■
Additional measured values
Definition of additional measured values of other measuring com-
mands which are saved as additional columns in the measuring point
list.