Metrohm tiamo 2.3 Patch 1 User Manual
Page 1414
6.7 Subwindow Sensors
■■■■■■■■■■■■■■■■■■■■■■
1398
■■■■■■■■
tiamo 2.3 Patch 1
Can only be edited if monitoring is switched on.
Default value
Set to work+ 999 days
Message
The message options can only be edited if monitoring is switched on.
Message by E-mail
on | off (Default value: off)
The message is additionally sent to the e-mail address defined under [E-
mail...] if this option is activated. The message is sent in text format.
[E-mail...]
[E-mail] opens the Send e-mail window (see Chapter 2.6.1, page 95).
Acoustic signal
on | off (Default value: off)
If this option is activated, an acoustic signal will be emitted additionally to
the message defined above.
Action
The actions can only be edited if monitoring is switched on.
If during sensor monitoring it is found that the working life has expired,
then one of the following actions will be triggered automatically at start
test:
Selection
Record message | Display message | Cancel
determination
Default value
Display message
Record message
The message that the working life of the sensor has expired will be
automatically saved with the determination.
Display message
A message is displayed and you can select whether you want to con-
tinue the run or cancel it. If the run is continued, then the message
that the working life of the sensor has expired will be saved automati-
cally with the determination.
Cancel determination
The running determination will be automatically canceled. The follow-
ing message must be confirmed with [OK].