Metrohm tiamo 2.3 Patch 1 User Manual
Page 1120
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1104
■■■■■■■■
tiamo 2.3 Patch 1
■
Calibration wavelengths
Input of the reference wavelengths. Definition of the behavior of the
calibration.
Command variables
The following command variables are generated in the method run by the
command CAL Spec and can be used in formulas under the designation
'Command name.Variable designation':
Designation
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
.C0
Coefficient c0 of the calibration curve
.C1
Coefficient c1 of the calibration curve
.C2
Coefficient c2 of the calibration curve
.C3
Coefficient c3 of the calibration curve
.CAL
Status of the calibration: 1 = calibrated normally, 0
= canceled, because reference wavelength was not
found.
.DBL
Total duration for the processing of the command in s
.STY
Type of stop with which the command was stopped: 1
= stopped normally, 0 = stopped manually with
[Quit] or [Stop] .
.CP{#}.MEA
Measured wavelength of the calibration point in nm. If
no measured peak can be assigned, then not found
will be entered.
.CP{#}.REF
Reference wavelength of the calibration point in nm.
.FIN
Command status: 1 = Command has been ended
at least once, 0 = Command was started but not
ended, invalid = Command was not started (vari-
able not available).
.TOL
Tolerance within which a measured peak can be
assigned to a calibration wavelength.