2 meas cond - general/hardware – Metrohm tiamo 2.3 Patch 1 User Manual
Page 1032
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1016
■■■■■■■■
tiamo 2.3 Patch 1
Designation
Description
.TC.C4
Coefficient c4 of the Chebyshev polynomial of the solu-
tion used
.TC.MAX
Maximum temperature coefficient in %/°C
.TC.MIN
Minimum temperature coefficient in %/°C
.TC.TSTART
Start temperature in °C
.TC.TSTOP
Stop temperature in °C
5.6.4.10.2
MEAS Cond - General/Hardware
Tab: Method
▶ MEAS Cond ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Input
25 characters
The general parameters for the device and the sensor are defined on this
tab.
Device
Device name
Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Display of the device type.
Selection
712 Conductometer | 856 Conductivity Mod-
ule
Default value
856 Conductivity Module
Sensor
Sensor
Selection of a sensor of the Conductivity sensor type from the sensors
configured in the sensor table.