6 det upol - potentiometric evaluation, Potentiometric evaluation – Metrohm tiamo 2.3 Patch 1 User Manual
Page 571
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1
■■■■■■■■
555
Titrando, 855
Range
0 ... 999999 s
Selection
off
Default value
off
Filling rate
Speed with which the dosing cylinder is to be refilled after the titration.
The maximum filling rate depends on the cylinder volume of the exchange
unit or dosing device used.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum
5.6.3.2.5.6
DET Upol - Potentiometric evaluation
Tab: Method
▶ DET Upol ▶ Properties... ▶ Potentiometric evaluation
Command name
Name of the command.
Input
25 characters
Parameters for the potentiometric evaluation of the titration curve with
automatic equivalence point recognition. One of the following three eval-
uation methods can be chosen here:
Potentiometric evaluation
Selection
Evaluation without window | Evaluation with
measured value window (Upol) | Evaluation
with volume window (mL)
Default value
Evaluation without window
Evaluation without window
With this option set, the two parameters EP criterion and EP recog-
nition will be applied across the entire range of the titration curve.