2 det upol - general/hardware, General/hardware – Metrohm tiamo 2.3 Patch 1 User Manual
Page 559
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1
■■■■■■■■
543
.SME
Start measured value (measured value after processing
the start conditions) in the unit of the measured value
.STE
Start temperature (temperature after processing of the
start conditions) in °C
.STY
Type of stop with which the command was stopped: 1
= normal; 0 = manual or after error
.SVA
Start volume absolute (volume that was added accord-
ing to the start condition "start volume") in mL
.SVM
Start volume measured value (volume which was
added according to the start measured value given in
the start conditions) in mL
.SVS
Start volume slope (volume that was added according
to the start condition "start slope") in mL
.SVT
Total start volume (volume that was added in depend-
ence on all three start conditions) in mL
.TITER
Titer value of the solution used for the command
5.6.3.2.5.2
DET Upol - General/Hardware
Tab: Method
▶ DET Upol ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Input
25 characters
The general parameters for the control device, the dosing device, the sen-
sor and the stirrer are defined on this tab.
Device
Device name
Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.