2 kfc - general/hardware – Metrohm tiamo 2.3 Patch 1 User Manual
Page 845
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1
■■■■■■■■
829
Designation
Description
.LP.TEM
Temperature for the last measuring point of the meas-
uring point list in °C
.LP.TIM
Time in s until the last measuring point in the measur-
ing point list is reached
.LP.UGE
Voltage at the generator electrode for last measuring
point of the measuring point list in V; in case of meas-
urements with 756, 831: 0 = not defined, 1 = <14
V, 2 = 14…28 V, 3 = >28 V
.MTE
Temperature measurement with sensor; 1 = on, 0 =
off
.NMP
Number of measuring points in measuring point list
.SME
Start measured value (measured value after processing
the start conditions) in the unit of the measured value
.STE
Start temperature (temperature after processing of the
start conditions) in °C
5.6.3.6.2
KFC - General/Hardware
Tab: Method
▶ KFC ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Input
25 characters
The general parameters for the device, the electrodes and the stirrer are
defined on this tab.
Device
Device name
Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.