5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
478
■■■■■■■■
tiamo 2.3 Patch 1
[Properties]
Open the dialog window in which the parameters for the selected win-
dow can be edited (see Chapter 5.6.3.2.2.10, page 484).
[Delete]
Delete the window selected in the table.
5.6.3.2.2.7
DET pH - Additional evaluations
Tab: Method
▶ DET pH ▶ Properties... ▶ Additional evaluations
Command name
Name of the command.
Input
25 characters
The following additional methods for evaluation of titration curves can be
activated and defined on this tab:
Fixed endpoint evaluation
Fix endpoint evaluation
on | off (Default value: off)
If this check box is activated, then the associated values for the other
measured quantities will be interpolated with a fixed measured quantity
(Measured value, Volume or Time) for the fix endpoint from the meas-
uring point list.
Moves the selected line up (changes the sequence).
Moves the selected line down (changes the sequence).
[New]
Open the dialog window Fixed endpoint evaluation # to enter a new
fixed endpoint (see Chapter 5.6.3.2.2.11, page 485).
[Properties]
Open the dialog window Fixed endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.2.2.11, page 485).
[Delete]
Delete the selected line.