Metrohm tiamo 2.3 Patch 1 User Manual
Page 478
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
462
■■■■■■■■
tiamo 2.3 Patch 1
Designation
Description
.ETE
End temperature (temperature following processing of
the command) in °C
.EVT
End volume (total dosed volume at the end of the
command) in mL
.FIN
Command status; 1 = Command has ended at
least once, 0, invalid (variable not available) = Com-
mand has never ended
.FP{x}.ERC
ERC for the fixed endpoint × (1…9)
.FP{x}.MEA
Measured value for the fixed endpoint × (1…9) in mV
.FP{x}.TEM
Temperature for the fixed endpoint × (1…9) in °C
.FP{x}.TIM
Time in s until the fixed endpoint × (1…9) is reached
.FP{x}.VOL
Volume for the fixed endpoint × (1…9) in mL
.HP{x}.MEA
Measured value for the HNP × (1…9) in mV (HNP =
half neutralization potential)
.HP{x}.TEM
Temperature for the HNP × (1…9) in °C
.HP{x}.TIM
Time in s until the HNP × (1…9) is reached
.HP{x}.VOL
Volume for the HNP × (1…9) in mL
.GP.VOL
Volume for the Gran endpoint in mL
.GP.MEA
Measured value for the Gran endpoint in the unit of
the measured value
.GP.TEM
Temperature for the Gran endpoint in °C
.GP.TIM
Time in s until the Gran endpoint is reached
.IME
Initial measured value (measured value before start
conditions are processed) in the unit of the measured
value
.ITE
Initial temperature (temperature before start conditions
are processed) in °C
.LP.CAx
Calculated value × (1...3) for the last measuring point
on the measuring point list
.LP.ERC
ERC for the last measuring point in the measuring point
list
.LP.EXx
External value × (1...3) for the last measuring point of
the measuring point list