3 meas t - measuring parameters, Measuring parameters – Metrohm tiamo 2.3 Patch 1 User Manual
Page 1007
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1
■■■■■■■■
991
Titrando, 855, 856, 867
Selection
1 | 2 | 3 | 4 | off
Default value
1
off
Means that no stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Titrando, 855, 856, 867
Range
-15 ... 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is activated, the stirrer will be switched off automatically
when the command has finished. This parameter is displayed only for
Titrandos, 855, 856 and 867.
5.6.4.7.3
MEAS T - Measuring parameters
Tab: Method
▶ MEAS T ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Input
25 characters
The parameters for the measuring procedure can be defined on this tab.
Measurement
The two measuring modes Measurement with drift control or Meas-
urement without drift control can be selected.
Selection
Measurement with drift control | Measure-
ment without drift control
Default value
Measurement with drift control
Measurement with drift control
The measurement is carried out drift-controlled if this option is ena-
bled. The measurement is stopped as soon as the defined Signal drift
or the Stop measured value is reached or a waiting time has
elapsed. The corresponding parameters are not visible if this option is
disabled.