6 data for specified electrode types – Metrohm tiamo 2.3 Patch 1 User Manual
Page 1358
6.3 Configuration data
■■■■■■■■■■■■■■■■■■■■■■
1342
■■■■■■■■
tiamo 2.3 Patch 1
after which the measured voltage has reached this comparison value
except for ±1 mV.
Range
0 ... 999 s
Default value
60 s
Min. offset potential
Minimum permitted offset voltage (voltage at pH = 7.0). This value applies
for all evaluations.
Range
-999 ... 999 mV
Default value
-15 mV
Max. offset potential
Maximum permitted offset voltage (voltage at pH = 7.0). This value
applies for all evaluations.
Range
-999 ... 999 mV
Default value
15 mV
6.3.3.6.6
Data for specified electrode types
Tab: Configuration
▶ Extras ▶ Templates ▶ Electrode types... ▶ [Proper-
ties]
▶ Very good electrode / Good electrode / Usable electrode
The three predefined electrode types Standard, Gel and Non-aqueous
are always present in the table and be neither edited nor deleted. The fol-
lowing limit values are defined for the evaluation of the electrodes of
these electrode types:
Electrode type "Standard"
Parameters
Excellent
electrode
Good
electrode
Usable
electrode
Max. streaming potential
(mV)
2.5
3.0
4.0
Max. drift (mV/min)
2.0
2.5
3.0
Min. slope (%)
96.5
96.0
95.0
Max. slope (%)
101.0
102.0
103.0
Max. response time (s)
45
50
60
Min. offset potential (mV)
-15
-15
-15
Max. offset potential (mV)
15
15
15