5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1200
■■■■■■■■
tiamo 2.3 Patch 1
Aspirate
Range
0.00000 ... 50.0 mL
Default value
10.0000 mL
Rate
Speed at which the commands Dosing, Fill, Aspirate, Eject and
Exchange position are carried out. The maximum rate depends on the
cylinder volume of the exchange unit or dosing device used. Aspiration
and Ejection of the sample should be done at a rate <10 mL/min.
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
5.6.6.5
PREP
5.6.6.5.1
PREP - Overview
Dialog window: Method
▶ PREP ▶ Properties... ▶ PREP - 'Command name'
Command for Rinsing cylinder and tubing of an exchange or dosing
unit.
Devices
This command can be executed with the following devices:
Titrando: 808, 809, 835, 836, 841, 842, 851, 852, 857, 859, 888, 890,
901, 902, 904, 905, 906, 907
Titrino: 736, 751, 758, 799
Dosing Interface: 846
Conductometer: 856
pH/Ion - Meter: 867
Sample Processor: 730, 774, 778, 789, 814, 815, 864, 874
Robotic Titrosampler: 855
Appearance
The command has the following appearance:
Parameters
The parameters for the command PREP are configured in the following
dialog window: