8 elt loop, 1 elt loop - overview, Elt loop – Metrohm tiamo 2.3 Patch 1 User Manual
Page 1131
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1
■■■■■■■■
1115
Min. waiting time
The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift contin-
ues to be checked while the waiting time is elapsing.
Range
0 ... 999999 s
Default value
10 s
Max. waiting time
If the signal drift has not yet been reached, then the measured value will
be accepted when the maximum waiting time has elapsed. If the waiting
time has not been newly entered then a waiting time that is suitable for
the drift will be calculated automatically according to the following equa-
tion:
Range
0 ... 999999 s
Default value
110 s
Measuring interval
Time interval for entering a measuring point in the measuring point list.
Range
0.1 ... 999999 s
Default value
2.0 s
Temperature
Temperature
The temperature at which calibration is carried out can be entered man-
ually. If a temperature sensor is connected and the Temperature meas-
urement is set on the General/Hardware tab under Sensor to auto-
matic or continuous, then the temperature will be measured continu-
ously. The value is used for temperature compensation with the pH meas-
urement.
Range
-20.0 ... 150.0 °C
Default value
25.0 °C
5.6.5.8
ELT LOOP
5.6.5.8.1
ELT LOOP - Overview
Dialog window: Method
▶ ELT LOOP ▶ Properties... ▶ ELT LOOP - 'Command
name'
Loop command for the electrode test (test of pH electrodes), used
together with the command ELT MEAS.