Metrohm tiamo 2.3 Patch 1 User Manual
Page 1074
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1058
■■■■■■■■
tiamo 2.3 Patch 1
Range
0.1 ... 999000 mAU/min
Default value
1.0 mAU/min
Min. waiting time
The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift contin-
ues to be checked while the waiting time is elapsing.
Range
0 ... 999999 s
Default value
0 s
Max. waiting time
If the signal drift has not yet been reached, then the measured value will
be accepted when the maximum waiting time has elapsed. If the waiting
time has not been newly entered then a waiting time that is suitable for
the drift will be calculated automatically according to the following equa-
tion:
Range
0 ... 999999 s
Default value
154 s
Measuring interval
Time interval for entering a measured value in the measuring point list.
Range
0.1 ... 999999.0 s
Default value
2.0 s
Stop measured value
Stop when the entered measured value has been reached since the start
of the measurement.
Range
0 ... 4000 mAU
Selection
off
off
off means that no stop will take place.
Measurement without drift control
Duration of measurement
Duration of the adsorption measurement.
Range
0 ... 999999 s
Default value
600 s
Measuring interval
Time interval for entering a measuring point in the measuring point list.