5 Method
tiamo 2.3 Patch 1
■■■■■■■■
1065
Measuring interval
Time interval for entering a measured value in the measuring point list.
Range
0.25 ... 999999.00 s (Increment: 0.25)
Default value
2.00 s
Stop measured value
Stop when the entered measured value has been reached since the start
of the measurement.
Range
0 ... 4000 mAU
Selection
off
Default value
off
off
off means that no stop will take place.
Temperature
Temperature
Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous,
then the temperature will be measured continuously.
Range
20.0 ... 60 °C
Default value
25.0 °C
Measurement without drift control
Duration of measurement
Duration of the adsorption measurement.
Range
0 ... 999999 s
Default value
120 s
Measuring interval
Time interval for entering a measuring point in the measuring point list.
Range
0.25 ... 999999.00 s (Increment: 0.25)
Default value
2.00 s
Stop measured value
Stop when the entered measured value has been reached since the start
of the measurement.
Range
0 ... 4000 mAU
Selection
off
Default value
off