Metrohm tiamo 2.3 Patch 1 User Manual
Page 1245
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1
■■■■■■■■
1229
Note
It is advisable to define an initial temperature if you work always at the
same temperature.
Heating time
The target temperature has to be reached within this time. The heating
rate can be modified with the heating time in order to program tempera-
ture ramps or entire temperature profiles (with several HEATER com-
mands). Off causes heating at the maximum possible heating rate (see
device manual).
This value is live editable for the 874
Range
1 ... 999 min
Selection
off
Default value
off
Max. waiting time
If the Target temperature is not reached within the defined Heating
time then this waiting time will start. This parameter is only visible if a
Heating time has been defined.
This value is live editable for the 874.
Range
1 ... 999 min
Selection
off
Default value
off
Action if timeout expires
on | off (Default value: on)
If this option is activated, then an action can be defined which is to be
carried out if the target temperature has not yet been reached after expiry
of the maximum waiting time (is displayed only when Timeout ≠ off).
Selection
Display message | Cancel determination | Can-
cel determination and series
Default value
Display message
Display message
A message is displayed and you can select whether you want to con-
tinue the run or cancel it.
Cancel determination
The determination is canceled and the next determination of the series
is started.