6 empty, 1 empty - overview, Empty – Metrohm tiamo 2.3 Patch 1 User Manual
Page 1219
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1
■■■■■■■■
1203
5.6.6.6
EMPTY
5.6.6.6.1
EMPTY - Overview
Dialog window: Method
▶ EMPTY ▶ Properties... ▶ EMPTY - 'Command name'
Command for Emptying cylinder and tubing of a dosing unit.
Devices
This command can be executed with the following devices:
Titrando: 808, 809, 835, 836, 841, 842, 851, 852, 857, 859, 888, 890,
901, 902, 904, 905, 906, 907
Titrino: 736, 751, 758, 799
Dosing Interface: 846
Conductometer: 856
pH/Ion - Meter: 867
Sample Processor: 730, 774, 778, 789, 814, 815, 864, 874
Robotic Titrosampler: 855
Appearance
The command has the following appearance:
Parameters
The parameters for the command EMPTY are configured in the following
dialog window:
■
Command variables
The following command variables are generated in the method run by the
command EMPTY and can be used in formulas under the designation
'Command name.Variable designation':
Identifi-
cation
Description
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command has
not ever been started
.CYL
Cylinder volume of the exchange or dosing unit used for
the command