5 Method
tiamo 2.3 Patch 1
■■■■■■■■
903
Titrino
Range
0.000 ... 9999.99 mL
Default value
100.00 mL
Selection
off
Stop time
Stop when the entered time has elapsed after the start conditions have
been fulfilled.
Range
0 ... 999999 s
Selection
off
Default value
off
since
Specifies when the stop time begins to be counted down (will not be dis-
played if the Stop time is disabled):
Selection
Start | Control point first reached | Last dosing
Default value
Start
Start
The stop time starts after the start volume has been dosed and after
the following pause time.
Control point first reached
The stop time starts as soon as the control point has been reached for
the first time.
Last dosing
The stop time starts after the last increment has been dosed, i. e. the
stop time is reset to zero again after each dosing step.
Stop rate
Rate at which the dosing cylinder is to be filled after titration. The maxi-
mum filling rate depends on the cylinder volume of the exchange unit or
dosing device used.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Selection
off
Default value
off
Titrino
Range
0.01 ... 150.00 mL/min
Selection
off
Default value
off
- 915 KF Ti-Touch (382 pages)
- 800 Dosino (53 pages)
- 767 Calibrated Reference (23 pages)
- 940 Professional IC Vario ONE/SeS/Prep 2 (54 pages)
- 754 Dialysis Unit (49 pages)
- 815 Robotic Soliprep for LC (76 pages)
- Vision Manual (207 pages)
- tiamo 2.1 Manual (1532 pages)
- 825 Lab Link (37 pages)
- 808 Titrando (70 pages)
- 902 Titrando (52 pages)
- 756 KF Coulometer (163 pages)
- 756 KF Coulometer (162 pages)
- 940 Professional IC Vario ONE/LPG (98 pages)
- 850 Professional IC Anion MCS Prep 3 (154 pages)
- 850 Professional IC Anion MCS Prep 3 (152 pages)
- 904 Titrando (58 pages)
- 850 Professional IC Anion MSM-HC MCS Prep 2 (150 pages)
- 930 Compact IC Flex Oven/ChS/Deg (47 pages)
- 872 Extension Module Liquid handling (64 pages)
- 814 USB Sample Processor (90 pages)
- 814 USB Sample Processor (91 pages)
- 940 Professional IC Vario (43 pages)
- Vision – Tutorial (40 pages)
- 799 GPT Titrino (242 pages)
- 889 IC Sample Center (68 pages)
- 761 Compact IC (228 pages)
- 851 Titrando (100 pages)
- 748 DH Sample Changer (32 pages)
- 940 Professional IC Vario ONE/SeS/HPG (51 pages)
- 896 Professional Detector – Amperometry (62 pages)
- 877 Titrino plus (139 pages)
- 881 Compact IC pro – Anion (129 pages)
- 940 Professional IC Vario ONE/ChS/HPG (112 pages)
- 930 Compact IC Flex Deg (41 pages)
- 840 PC Control 5.0 / Touch Control (351 pages)
- 940 Professional IC Vario ONE/Prep 1 (45 pages)
- 776 Dosimat (42 pages)
- 717 Sample Changer (36 pages)
- 815 Robotic USB Sample Processor XL (113 pages)
- 815 Robotic USB Sample Processor XL (114 pages)
- 940 Professional IC Vario ONE/SeS/PP (126 pages)
- 838 Advanced Sample Processor Installation Instructions (109 pages)
- 700 Dosino (55 pages)
- 719 S Titrino (152 pages)