Metrohm tiamo 2.3 Patch 1 User Manual
Page 1065
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.3 Patch 1
■■■■■■■■
1049
Minimum evaluation
Minimum evaluation
on | off (Default value: off)
If this option is activated, then the associated time for the minimum meas-
ured value will be interpolated from the measuring point list.
Threshold value
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.
Range
0.1 ... 1000.0 mAU/s
Default value
20.0 mAU/s
Maximum evaluation
Maximum evaluation
on | off (Default value: off)
If this option is activated, then the associated time for the maximum
measured value will be interpolated from the measuring point list.
Threshold value
The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.
Range
0.1 ... 1000.0 mAU/s
Default value
20.0 mAU/s
Break point evaluation
Break point evaluation
on | off (Default value: off)
If this option is activated, then sharp, almost 90° changes in directions in
the measurement curve will be evaluated.
EP criterion
Measure of the minimum sharpness of the break point. The smaller the EP
criterion set, the more break points will be found. As this is a relative value
related to the total measured value modification, even small changes in
the measured value can be evaluated as a break point for a small meas-
ured value range.
Range
0.0 ... 1.0
Default value
0.3